The hCG One Step Pregnancy Test Strip (Urine) detects hCG at a concentration of 25mIU/mL or greater. The test has been standardized to the W.H.O. International Standard. The addition of LH (500mIU/mL), FSH (1,000mIU/mL), and TSH (1,000µIU/mL) to negative (0mIU/mL hCG) and positive (25mIU/mL hCG) specimens showed no cross-reactivity.
The following potentially interfering substances were added to hCG negative and positive specimens.
Acetaminophen | 20 mg/dL | Caffeine | 20 mg/dL |
Acetylsalicylic Acid | 20 mg/dL | Gentisic Acid | 20 mg/dL |
Ascorbic Acid | 20 mg/dL | Glucose | 2 g/dL |
Atropine | 20 mg/dL | Hemoglobin | 1 mg/dL |
Bilirubin | 2 mg/dL |
None of the substances at the concentration tested interfered in the assay.
READING THE RESULTS
PREGNANT
Two distinct colored lines appear. One line should be in the control line region (C) and another line should be in the test line region (T). One line may be lighter than the other; they do not have to match. This means that you are probably pregnant.
NOT PREGNANT
One colored line appears in the control line region (C). No line appears in the test line region (T). This means that you are probably not pregnant.
INVALID
The result is invalid if no colored line appears in the control line region (C), even if a line appears in the test line region (T). You should repeat the test with a new test strip.
1.Professional Manufacturer, a national-level technologically advanced “giant” enterprise
2.Deliver goods as order request by sea, by air, by express
3.ISO13485, CE, Prepare various shipping documents
4.Reply to client inquiries within a day