Medical use professional Typhoid Test kit, one step rapid test cassette

Short Description:


Product Detail

Product Tags

Clinical Sensitivity, Specificity and Accuracy

The Influenza A+B Antigen Rapid Test has been tested compared with RT-PCR. 539 nasopharyngeal swabs and oropharyngeal swabs were evaluated with the Influenza A+B Rapid Test.

Substances

Concentration

Substances

Concentration

Nasal Spray

15% v/v

Hemoglobin

10% v/v

Mucin

0.5 % w/v

Mupirocin

10 mg/mL

Nasal Drops

15% v/v

Mouthwash

/

Chloraseptic

1.5 mg/mL

Levofloxacin

40 ug/mL

 Oseltamivir

2ug/mL

Ribavirin

0.2ug/mL

Fluticasone propionate

5% v/v

Ceftriaxone

800 ug/mL

Tobramycin

4ug/mL

Saline Nasal Spray

10% v/v

For Influenza A

Method

RT-PCR

Total Results

Influenza A+B Rapid Test

Results

Positive

Negative

Positive

116

1

117

Negative

5

417

422

Total Results

121

418

539

Clinical sensitivity:95.87% (95% CI: 90.69%~98.22%)
Clinical specificity:99.76% (95% CI: 98.66%~99.96%)
Total Coincidence rate:98.89% (95% CI: 97.59%~99.49%).

For Influenza B:

Method

RT-PCR

Total Results

Influenza A+B Rapid Test

Results

Positive

Negative

Positive

97

1

98

Negative

6

435

441

Total Results

103

436

539

Clinical sensitivity:94.17% (95% CI: 87.87%~97.30%)
Clinical specificity:99.77% (95% CI: 98.71%~99.96%)
Total Coincidence rate:98.70% (95% CI: 97.34%~99.37%).

Analytical Sensitivity/LOD

 img-1 Hangzhou Aichek Medical Technology CO., Ltd.

Jinxing Cun, Yuhang Community, Yuhang

District (Future Sci-Tech City), Hangzhou,

Zhejiang, P.R. China

 img-2 SUNGO Europe B.V.

Olympisch Stadion 24, 1076DE Amsterdam, Netherlands

The limit of detection (LOD) was identified by evaluating different concentrations of influenza A virus and influenza B virus in the Influenza A+B Antigen Rapid Test. The concentrations identified as the LOD levels tested are listed below.
Influenza A (H3N2): 5×103 TCID50/mL
Influenza A (H1N1): 2.5×103 TCID50/mL
Influenza A (H1N1 pdm09): 2.5×103 TCID50/mL
Influenza B (Yamagata): 3.5×103 TCID50/mL
Influenza B (Victoria): 1.0×103 TCID50/mL

Analytical Specificity (Cross Reactivity)

To determine the analytical specificity of the Influenza A+B Antigen Rapid Test, several commensal or pathogenic microorganisms that may be present in the upper respiratory tract were tested.
Positive and negative specimens were spiked with these microbes were evaluated at a concentration of 106 TCID50/mL, including SARS-CoV-2, Human coronavirus HKU1, OC43, NL63, 229E, MERS, Rhinovirus, Adenovirus, Enterovirus, Metapneumovirus, Parainfluenza, Respiratory syncytial virus, Mycoplasma pneumonia, Chlamydia pneumonia, Streptococcus pneumonia, Staphylococcus aureus, Mycobacterium tuberculosis, Haemophilus influenzae, Streptococcus pyogenes. No cross-reactivity was seen with the Influenza A+B Antigen Rapid Test.

INTENDED USE

Influenza A+B Antigen Rapid Test is a lateral flow immunoassay intended for the qualitative detection influenza A and influenza B antigens in nasopharyngeal swab and oropharyngeal swab.